thickness with a constant of 525 would indicate an applied voltage of 2,100 volts.  A 

common practice used in setting inspection voltages in the field is to adjust the output 

voltage by visual observation.  It is the general consensus that spark discharge at 

least twice the thickness of the coating will give adequate inspection voltage and 

compensate for any irregularity in coating thickness and grounding conditions.  If this 

practice is desired for determining inspection voltage, it should be done while the 

electrode is in the normal operating position and under actual grounding conditions. 

 

The formulas and suggestions for setting voltage are supplied as a guide, and are not 

necessarily recommended by the manufacturer of this detector.  

 

Consult the coating manufacturer for recommended voltage applications. 

 

5. 

INSTRUMENT SERVICING INSTRUCTIONS

 

 

 

 

A.  Cleaning:  Keep the instrument clean and dry.  Clean the instrument case with a 

soft cloth dampened with kerosene, then wipe dry.  Do not use solvents such as 

lacquer thinner, methyl ethyl keytone, etc. 

 

KEEP ALL ELECTRICAL CONTACTS CLEAN 

B.  Voltage Output Checks 

A.  In case of LOW output voltage: 

a. 

Check position of voltage selector switches. 

b. 

Check for “parted” conductor in voltage probe and/or ground trail. 

B.  In case of NO output voltage: 

a. 

Check power cord and plug. 

b. 

Check for “parted” conductor in voltage probe and/or ground trail. 

 

 

6. 

FACTORY REPAIRS

 

 

Holiday Detectors returned to the factory for repairs should be sent 

TRANSPORTATION PREPAID.  In most cases the detector can be repaired and 

returned the same day it is received at the factory

 

WHEN ORDERING PARTS FOR YOUR DETECTOR OR REQUESTING FURTHER INFORMATION 

ALWAYS GIVE THE DETECTOR’S SERIAL NUMBER.

 

 

 