![Time Electronics 7085A User Manual Download Page 13](http://html1.mh-extra.com/html/time-electronics/7085a/7085a_user-manual_1116926013.webp)
Time Electronics
User Manual
7085A Temperature Distribution and 10 Channel Scanner Module
v1.1
7085A Temperature Distribution and 10 Channel Scanner Module
Page 13 of 21
2.6.2.4
Test Details
From the wizard a sequence of tests is created. Normally this is all that is necessary, but
each test can be individually edited if required.
The example above shows a series of tests using a reference pt100 probe to calibrate 5
thermocouples from 25 to 125 °C.
The procedure is created using a ‘Reference Reading’ for each test point. Each UUT reading
is then compared against the reference reading test using the auxiliary reading function.
Within the auxiliary reading function the input method is set to use ‘Result from previous test’
and the test number option is selected. This is the test result to compare against.
Note:
If any tests are deleted or added the ‘result from previous test’ will be out of sequence. In this case the test
numbers must be amended manually.