7085A Temperature Distribution and 10 Channel Scanner Module
v1.1
7085A Temperature Distribution and 10 Channel Scanner Module
Page 12 of 21
2.6.2.2
Output
This is the unit under test. The output range is usually the equal to the input range. The
Calibrating Instrument is the DMM used to measure the output of the thermocouple or other
type of temperature sensor.
CJC & Thermocouple Type
For thermocouple sensors the CJC reference and thermocouple type options are available. In
most cases the CJC reference is set to 7085.
Use Scanner
Tick this option to utilise the automatic switching capabilities of the 7085A
2.6.2.3
Details
Number of devices under test
This determines the number of channels which will be used. A maximum of 8 thermocouples
can be calibrated using the 7085A.
Number of test points
The range is divided by the number of test points.
Method
For temperature sensors this is usually up. The procedure wizard will create test points from
0 to 100% of the range entered.