22
the horizontal pushing motor and the motor synchronous pulley is loose.
4. If all items above are normal, replace the main control panel. You need to enter the commissioning mode
before replacing the main control panel to record all motor parameters. After the replacement, manually
enter all parameters and check the alignment position of the instrument.
Maintenance Methods:
1. If the shading plate and photoelectric switch cannot sense each other, re-adjust the position of the
shading plate and place the shading plate at the central position of 2 corners of the photoelectric switch.
1. If the motor does not run and the red indicator of the photoelectric switch is off, you need to replace
with new materials, remove the back panel, replace the faulty components with new ones, and re-route
the wires according to the original wiring position, as shown in the figure below.
3. If there is a small amount of displacement in the experiment module, it means that the connection
between the horizontal pushing motor and the motor synchronous belt pulley is loose. Remove the
motor base from the bottom plate and re-lock the motor synchronous belt pulley as follows:
Install two M4*8 or M3*6 (according to the actual conditions) jackscrews in the fixed holes of the
synchronous belt pulley of the horizontal pushing motor. One jackscrew must be pressed against the
Horizontal pushing
photoelectric switch
Horizontal pushing
shading plate
Horizontal
pushing motor