39
man_vd6s2-de-130515
In order to adjust the switchpoints:
Press the menu key (several times), until the LED "setpoint 3-state-control" is on.
Adjust the setpoint by means of the "Up" and "Down" keys.
Confirm with "Set" key. The display shows the current hysteresis H1 for evacua-
tion.
Adjust the hysteresis H1 by means of the "Up" and "Down" keys. The
maximum permissible hysteresis is 90% of the setpoint.
Confirm with "Set" key. The display shows the current hysteresis H2 for venting.
Adjust the hysteresis H2 by means of the "Up" and "Down" keys. The
maximum permissible hysteresis is 90% of the setpoint.
Confirm with "Set" key.
set
set
set