23
3.3.2 Operation with other Supply and Evaluation Units
The transducer can be operated with other customer related display units or volta-
ge supplies.
The electrical connection is to be made by means of suitable cables considering
EMI demands and according to the pin description shown below:
Socket
M12, A-coded, 5pole, male
4
3
2
5
1
Pin1:
Signal Output 4-20 mA
Pin2:
Do not connect
Pin3:
Voltage Supply 12...30 VDC
Pin4:
Supply GND
Pin5:
Shield
Maximum admissible load resistor in
Ω
:
(Supply Voltage[V] - 4 V) / 0.02 A
Incorrect connection or inadmissible supply voltage can damage
the transducer!
VSR-4-20 mA-210805