MyoTrac Infiniti
™
User Guide
35
Edit Wizard
The edit wizard guides you through each
of the parameters that are available to
change in the software
– Work, Rest
durations, Ramp Time, Pulse Rate and
Pulse Width. Note that for Incontinence
stim sessions, Pulse Rate and Pulse
Width cannot be modified. Once the
changes have been made, review them in
the Stim settings overview before
continuing. Stim intensity is controlled
from the stimulation screen.
Stim Delivery Screen
The Stim delivery screen is the screen that
displays the current intensity, maximum intensity
and its controls for each of the two channels.
Tap the Start button followed by the up arrows to
incrementally control the current being delivered.
Current can only be increased during the work
period, not during the ramp times or during the
rest period. The current can be decreased during
the work or rest periods, but not the ramp times.
The maximum current set for the work period is
displayed for each channel in the diagram in the
middle of the screen. The actual supplied current
is shown as the larger numbers at the top of the
control arrows. This supplied current reading will
display the max mA during work, 0mA during rest
and an increasing and deceasing value during the
ramps. If you keep your finger on the control
buttons it will continue to increase until it reaches
the maximum of 100mA.
If at anytime you wish to stop the session press
the
STOP
button that covers the top portion of the
screen. As a safety feature if the MyoTrac Infiniti
detects the electrodes being removed while the
unit is stimulating, it will automatically stop the
Stim. The illustration of the stim cycle in the
middle of the screen indicates along with the label
at the top of the screen the period of the stim you
are currently in.
Displayed in the status bar of the stimulation
screen is the number of sets done, for an open
stimulation setting, or the number remaining for a
scripted session.