© 2016 Thorlabs
28
TED8000
6 Communication with a PC
The description of the PRO8000 Series mainframe includes all instructions of how to prepare and
execute the programming of the system via a computer interface.
Special operation features of the TED8000 module are described here. See also section
.
Note
All analog values are read and written in SI units, i.e. A (not mA), W (not mW) etc. Letters may be
written in lower or upper cases.
Attention
Prior to programming a TED8000 module the limit value of the TEC current I
LIM
(hardware limit) for
the used TEC element must be set using a screwdriver.
The corresponding potentiometer is marked I
LIM
and is located on the
of the
TED8000 module.
The value I
LIM
is constantly measured by the PRO8000 Series mainframe and can be checked in
of the TED8000 during setting.
6.1 Nomenclature
Program messages (PC to PRO8000) are written in inverted commas:
"*IDN?"
Response messages (PRO8000 to PC) are written in brackets:
[
:SLOT 1
]
Decimal point:
1.234
Subsequent parameters are separated with commas:
"PLOT 2,0"
Subsequent commands are separated with semicolons:
*IDN?;*STB?"
6.2 Data Format
According to the IEEE 488.2 specifications all data variables are divided into 4 different data
formats:
Character response data (<CRD>)
is a single character or a string.
Examples:
A
or
ABGRS
or
A125TG
or
A1.23456A
, section 8.7.1)
Numeric response data Type 1 (<NR1>)
is a numerical value with sign in integer notation.
Examples:
1
or
+1
or
-22
or
14356789432
, section 8.7.2)
Numeric response data Type 2 (<NR2>)
is a numerical value with or without sign in floating point
notation without exponent.
Examples:
1.1
or
+1.1
or
-22.1
or
14356.789432
, section 8.7.3)
Numeric response data Type 3 (<NR3>)
is a numerical value with or without sign in floating point
notation with exponent with sign.
Examples:
1.1E+1
or
+1.1E-1
or
-22.1E+1
or
143.56306
, section 8.7.4)
17
8
17
50
50
50
50
Summary of Contents for TED8000 Series
Page 1: ...PRO800 PRO8000 4 Series Modular Temperature Controllers TED8000 Operation Manual 2016 ...
Page 2: ...Version Date Item No 3 2 08 Sep 2016 M0009 510 227 Copyright 2016 Thorlabs ...
Page 42: ... 2016 Thorlabs 40 TED8000 Structure of the registers DEC DEE and EDE ...
Page 43: ... 2016 Thorlabs 6 Communication with a PC 41 Structure of the registers ESR ESE STB and SRE ...
Page 59: ......