© 2016 Thorlabs
26
TED8000
5.3.9 Setting the P, I and D Share of the Control Loop
The temperature control behavior of the TED8000 can be adapted to the individual laser setup by
optimizing the P, I and D share parameters of the control loop.
They can be set separately in a range between 0.1 % and 100 %:
Psh
=
P share
Ish
=
I share
Dsh
=
D share
To change the three parameters select them one by one and set them to the desired value.
Press
to apply the setting.
For adjustment of the parameters, it is sometimes necessary to switch off the I-share completely.
There is a separate switch parameter for this purpose:
Ishare = ON/OFF
Toggle the function using the right soft key.
Press
to terminate input.
5.4 Switching ON and OFF
Attention
TED8000 modules can be switched on or off at any time, with no regard if any parameters have
been set! So make sure that the appropriate
are made prior to switching on the
module!
Select the module to be switched on or off in the
. The LED “SEL” of the selected
module lights up.
Press the key
to switch on the selected module. The LED “ON” of the selected module will
light up, this way indicating that the TEC current is enabled.
17
18
Summary of Contents for TED8000 Series
Page 1: ...PRO800 PRO8000 4 Series Modular Temperature Controllers TED8000 Operation Manual 2016 ...
Page 2: ...Version Date Item No 3 2 08 Sep 2016 M0009 510 227 Copyright 2016 Thorlabs ...
Page 42: ... 2016 Thorlabs 40 TED8000 Structure of the registers DEC DEE and EDE ...
Page 43: ... 2016 Thorlabs 6 Communication with a PC 41 Structure of the registers ESR ESE STB and SRE ...
Page 59: ......
Page 60: ...www thorlabs com ...