Page 35
16005-D03
T-Cube TEC Controller
Appendix A Connector Pinout Details
A.1 TEC OUTPUT Connector
The TEC OUTPUT connector on the rear panel provides connection to the Laser
Diode Mount or other device. If Laser Diode Mounts by Thorlabs are used, connect
the "TEC OUTPUT" of the T TC001 to the 9-pin plug "TEC DRIVER" of the Laser
Diode Mount, using the provided shielded cable CAB420-15.
With other equipment, connect the TEC element and the temperature sensor as
detailed in Fig. A.1 below.
Fig. A.1 TEC OUTPUT Connector Pin Identification
Pin
Description
Pin
Description
1
Not Used
9
Not Used
2
Not Used
10
Transducer AD592 (-ve)
3
Thermistor -ve, ground
11
Transducer AD592 (+ve)
4
Therve
12
Not Used
5
TEC (+)
13
TEC (-)
6
TEC (+)
14
TEC (-)
7
TEC (+)
15
TEC (-)
8
Not Used
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15