See below for possible troubleshooting issues. 

Problem 

Solutions 

Power LED not enabled 
while pressing the Power ON 
button 

Check the AC cord IEC connector is properly 

connected to the switching power supply and that it 
is connected to a live AC mains outlet. 

Check DC plug is fully inserted into the SLSLLG2 DC 

jack labeled; 5 VDC 2.4 A. 

Shutter does not open or 
close 

Check the SLSLLG2 shutter cable 4 pin Mini-Din plug 

has arrow facing up and is fully inserted into the 
connector labeled OUTPUT. 

Status LED turned amber.  

In Protection Mode; Status LED should be green 

under normal operating conditions, and turn amber 
as a warning to state there is 3 minutes until 
shutdown (at modulation frequencies of 11 to 
15 Hz). See Section 4.5, page 10. 

Status LED turned Red. 

In Protection Mode: Based on modulation frequency if 

>10 Hz a timeout period is applied to disable the 
shutter drive circuit after the amber LED 3 minute 
warning. 

Shutter operation is erratic 
or stuck 

Verify the shutter head is not located near magnetic 

sources or directly against ferrous materials, 
especially the front surface. 

Unit powers on when DC 
power is applied 

Default to On

” mode has been set. Follow steps on 

page 11 to disable. 

 

CAUTION 

Prolonged modulation of these shutters will eventually result in premature 

component wear and failure. This device is not intended for long-term 

continuous modulation. 

!

!

Summary of Contents for SLSLLG2

Page 1: ...SLSLLG2 Liquid Light Guide Adapter with Shutter User Guide...

Page 2: ...djusting the Position of the LLG Adapter 7 4 3 Manual Shutter Operation 8 4 4 External Modulation Input 9 4 5 Protected Mode 10 4 6 Unprotected Mode 10 4 7 Default to On 11 4 8 Interlock Output 11 4 9...

Page 3: ...quid Light Guide Adapter with Shutter 9 1 LLG Adapter and Shutter 20 9 2 Shutter Controller 21 Chapter 10 Regulatory 22 Chapter 11 Certificate of Conformance 23 Chapter 12 Thorlabs Worldwide Contacts...

Page 4: ...ce Symbol Description Direct Current Alternating Current Both Direct and Alternating Current Earth Ground Terminal Protective Conductor Terminal Frame or Chassis Terminal Equipotentiality On Supply Of...

Page 5: ...ere are no user serviceable parts in this product CAUTION DO NOT OPEN HOUSING The SLSLLG2 shutter and controller has no user serviceable parts Do not operate without cover installed WARNING Do not ope...

Page 6: ...on will open and close the shutter To use the modulation input press and hold the ENABLE button for 3 seconds The SLSLLG2 will power on in the last used mode either Manual or External Modulation Mode...

Page 7: ...Liquid Light Guide Adapter with Shutter Chapter 3 Description Page 4 CTN008800 D02 3 1 Overview Figure 1 SLSLLG2 Side View Figure 2 SLSLLG2 Mounted on SLS401 Light Source...

Page 8: ...er Chapter 3 Description Rev A September 8 2016 Page 5 Figure 3 Controller Front View Figure 4 Controller Rear View Enable Button Status LED Power LED Power Button Interlock Jack External Modulation I...

Page 9: ...AC DC switching power supply rated at 100 240 VAC 47 63 Hz and 5 VDC at 2 4 A maximum output 4 Place the controller on a flat surface Connect the mini DIN connector from the shutter head with the arro...

Page 10: ...id light guide adaptor Item AD3LLG or AD5LLG to get the strongest light from the light source The adjustment can be done as shown in below 1 Use a spanner wrench Item SPW602 to loosen the retaining ri...

Page 11: ...ning of the LED colors and modes please see the table on page 8 and page 12 Condition Shutter State Open Closed Depressing the ENABLE button Closes Opens Power plugged into either wall outlet or contr...

Page 12: ...ted TTL signal connect the signal source to the shutter controller at the BNC TTL Input connector Activate the TTL signal source The shutter will be in the last state it was in when powered on therefo...

Page 13: ...n this mode the unit will disable the blades turn the Status LED red and trip the interlock when the unit is run over 10 Hz and the time limit is reached see Specifications on page 15 for more informa...

Page 14: ...act Thorlabs Tech Support if you believe your unit may need to be updated to include this feature 4 8 Interlock Output An interlock output is provided In the event of a failure the interlock circuit w...

Page 15: ...n the mode When the Status LED turns red the device will shut down until the user cycles the power However when the device is in Unprotected Mode the Status LED will still turn amber 15 seconds before...

Page 16: ...ully opening D E 15 17 Delay from fall of input pulse to initiation of shutter fully closing E G 9 11 Falling edge to 100 closed D G 24 26 Delay from fall of input drive to shutter fully closed Minimu...

Page 17: ...delay from the initiation of the starting pulse to the shutter responding IA is typically 11 ms Similarly the closing time DE is typically 15 ms The mechanical rise time of the shutter is given as AC...

Page 18: ...ecifications Specification Value Lens ACL5040U A Glass Type B270 Optical Crown Glass Wavelength Range 350 700 nm Effective Focal Length 40 mm Back Focal Length 26 mm Clear Aperture 45 mm Typical Coupl...

Page 19: ...Hz 1 7 ms Typ 4 ms Max Jitter Falling Edge 10 Hz 1 7 ms Typ 4 ms Max Jitter Rising Edge 11 15 Hz 4 ms Typ 8 ms Max Jitter Falling Edge 11 15 Hz 4 ms Typ 8 ms Max Rise Time 10 ms Typ 12 ms Max Fall Tim...

Page 20: ...e connector on the SLSLLG2 is a 4 pin Mini DIN shielded female connector that accepts the 4 pin Min Din male connector The pin descriptions are as follows Pin Description 1 Shutter Detect 2 Modulation...

Page 21: ...ades to rust The SLSLLG2 use five blades for the shutter These blades are thin 0 003 and lightweight making them very delicate Please do not insert any objects into the shutter where the blades may be...

Page 22: ...Protection Mode Status LED should be green under normal operating conditions and turn amber as a warning to state there is 3 minutes until shutdown at modulation frequencies of 11 to 15 Hz See Section...

Page 23: ...Adapter and Shutter 62 0 mm 2 4 79 3 mm 3 12 61 9 mm 2 44 Cable Length 889 mm 1 0 mm of Adjustment Loosen Retaining Ring to Adjust 26 5 mm 1 04 SM2 2 035 40 Series External Thread 16 0 mm 0 63 Retaini...

Page 24: ...Liquid Light Guide Adapter with Shutter Chapter 9 Mechanical Drawings Rev A September 8 2016 Page 21 9 2 Shutter Controller 3 00 76 2mm 3 42 86 7mm 3 50 88 9mm 1 38 35 1mm...

Page 25: ...other Thorlabs products such as Pure OEM products that means assemblies to be built into a unit by the user e g OEM laser driver cards Components Mechanics and optics Left over parts of units disasse...

Page 26: ...Liquid Light Guide Adapter with Shutter Chapter 11 Certificate of Conformance Rev A September 8 2016 Page 23 Chapter 11 Certificate of Conformance...

Page 27: ...ormation USA Canada and South America Thorlabs Inc sales thorlabs com techsupport thorlabs com Europe Thorlabs GmbH europe thorlabs com France Thorlabs SAS sales fr thorlabs com Japan Thorlabs Japan I...

Page 28: ...www thorlabs com...

Reviews: