TIME CLOSED (OFF)
When the active window indicator is in
TIME CLOSED
quadrant (lower left) the
▲
▼
keys will increment
and decrement the indicated digit. The
◄►
keys will move between digits. This indicates how long the
shutter will be closed (when the shutter is not connected it indicates the “low” time of the TTL output
trigger).
NOTE
: If active window indicator is in the
TIME CLOSED
quadrant when the shutter is enabled, timing will begin
with the programmed
TIME CLOSED
.
INTERLOCK and ALARM
Using the Safety Interlock - Connections and Operation
The SC10 is equipped with a Remote Interlock connector located on the rear panel (Refer to Figure 1). In order
to enable the controller, a short circuit must be applied across the terminals of the Remote Interlock connector. In
practice this connection is made available to allow the user to connect a remote actuated switch to the connector
(i.e. an open door indicator). The switch (which must be normally open) has to be closed in order for the unit to be
enabled. Once the switch is in an open state and an SH05 is connected, the shutter will automatically close. And
the display will flash “
INTERLOCK OPEN
”. If the switch returns to a closed condition the shutter must be re-
enabled at the unit by pressing the
ENABLE
switch.
All units shipped from Thorlabs are configured with a shorting device installed in the Interlock connector. If you
are not going to use this feature then you can leave the shorting device installed and the unit will operate normally
as described in the procedures above.
If you wish to make use of the Interlock feature you will need to acquire the appropriate connector mate and wire
it your remote interlock switch. Next, remove the shorting device by unscrewing it from the input and install the
connector into the Interlock input.
The Interlock input only accepts a
2.5mm mono phono jack
. This connector is readily available at most
electronics stores (Radio Shack, Digikey, Mouser, Allied to name a few).
The electrical specifications for the Interlock input are as follows:
Type of Mating Connector: 2.5mm mono phono jack
Open Circuit Voltage: +5VDC with respect to Chassis ground
Short Circuit Current: 0.5mADC
Connector Polarity: Tip is +5V, Barrel is ground
Interlock Switch Requirements: Must be N.O. dry contacts (under no circumstances should any external
voltages be applied to the Interlock input)
8654-D02 Rev. B 8/15/05
Page 7 of 12
Figure 2
- Remote Interlock Connector
ALARM
The SC10 has an Alarm feature when used in conjunction with an SH05 shutter. In the event that the shutter does
not close within 30 ms of the deactivation voltage (Shutter Closed State) the display will indicate an “ALARM”
condition. This feature is only active when the SH05 shutter is connected.