© 2016 Thorlabs GmbH
8 Appendix
61
8 Appendix
8.1 Technical Data
Current Inputs
Full Scale Measurement Ranges
100 nA … 10 mA (in decade steps)
Maximum Display Resolution
10 pA
Photodiode Polarity
cathode grounded (CG) or anode grounded (AG)
Display Units
A, W, dBm
Bias Voltage
0 … -10 V (CG), 0 .. +10 V (AG)
Max. Photodiode Capacitance
10 nF
Input Impedance
~0
W
(virtual ground)
Current Range
Resolution
Gain
Accuracy
Bandwidth
10 mA
1 µA
1 x 10² V/A
+/- 0.2% f.s.
100 kHz / 25 Hz
1 mA
100 nA
1 x 10³ V/A
+/- 0.2% f.s.
100 kHz / 25 Hz
100 µA
10 nA
1 x 10
4
V/A
+/- 0.2% f.s.
100 kHz / 25 Hz
10 µA
1 nA
1 x 10
5
V/A
+/- 0.2% f.s.
100 kHz / 25 Hz
1 µA
100 pA
1 x 10
6
V/A
+/- 0.2% f.s.
10 kHz / 25 Hz
100 nA
10 pA
1 x 10
7
V/A
+/- 0.5% f.s.
1 kHz / 25 Hz
Voltage Input Thermal Sensors
Full Scale Measurement Ranges
1 mV … 1 V (in decade steps)
Maximum Display Resolution
0.1 µV
Display Units
V, W, dBm
Input Impedance
1 M
W
Time constant correction
< 1 s to 30 s
Voltage Range
Resolution
Gain
Accuracy
Bandwidth
1 V
100 µV
1 x 10
0
+/- 0.5% f.s.
1.5 Hz
100 mV
10 µV
1 x 10
1
+/- 0.5% f.s.
15 Hz
10 mV
1 µV
1 x 10
2
+/- 0.5% f.s.
7 Hz
1 mV
0.1 µV
1 x 10
3
+/- 0.5% f.s.
0.7 Hz
Voltage Input Pyroelectric Sensors
Full Scale Measurement Ranges
100 mV … 100 V (in decade steps)
Maximum Display Resolution
10 µV
Display Units
V, J, W, J/cm², W/cm²
Input Impedance
1 M
W
Measurement Repetition Rate
0.1 Hz to 1500 Hz
Voltage Range
Resolution
Gain
Accuracy
Rep. Rate
100 V
10 mV
1 x 10
-2
+/- 0.5% f.s.
1.5 kHz
10 V
1 mV
1 x 10
-1
+/- 0.5% f.s.
1.5 kHz
1 V
100 µV
1 x 10
0
+/- 0.5% f.s.
1.5 kHz
100 mV
10 µV
1 x 10
1
+/- 0.5% f.s.
1.5 kHz