3 Operating Instructions

Rev: 1.0,  15-Mar-2021

Page  14

3.2.14   NTC Input

To monitor the temperature in a test environment, the PM103 has an NTC Input (DA-15 Pin 14)
which allows connection of an NTC thermistor. The measurement range is 0.1 – 100 k

W

. The

set range R0 and the bandwidth B for the NTC thermistor can be adjusted in the OPM software
or through SCPI commands. The default settings are 10 k

W

 and 3988 K to directly access the

TSP-TH

 temperature probe. The measurement range with the PM103 is then -10 °C to 80 °

C with the TSP-TH.

 Note 

Connection of the NTC to the DE-9 Pin 7 of the Sensor Connector requires a resistance of 0.2
k

W

 to 40 k

W

.

3.2.15   Power Supply

All PM103x models use the USB port for power supply. Please provide an appropriate connec-
tion to a PC or power bank. 

 Note 

When operating the PM103 via the RS232 interface or completely autonomously via an analog
output, the unit can either be powered with 5 V via the USB connector or with any supply in the
range  from  +5  V  to  +36  V  via  the  DA-15  connector  (Pin  1  and  Pin  9  (GND)).  As  an  external
power supply, we recommend the 

DS5

 5 VDC regulated power supply.

3.3   Software

The PM103x is operated remotely from a PC, laptop or tablet.

Thorlabs OPM Software

The  Thorlabs 

Optical  Power  Monitor

  (OPM)  software  provides  a  Graphical  User  Interface  to

view, log, and analyze measurement data. Download the OPM software package Version 2.3 or
higher from the Thorlabs website, via this 

link

, through the software tab on the product website,

or through searching for the product name within the Thorlabs 

software website

.

All information on software installation, device control and data  management  software  can  be
found in the software manual directly downloaded 

here

.

 Note 

Do not connect the PM103x prior to software installation!

 Attention 

Exit all running applications on your PC as the installer may require a reboot of your PC during
installation! 

Custom Made Software

The PM103x can also be run using custom made software. For this, please refer to the docu-
ment Write-Your-Own-Application (WYOA) for PM103x.

14

Summary of Contents for PM10-3

Page 1: ...Optical Power Meter Interfaces 2021 PM103 PM103A PM103U Operation Manual...

Page 2: ...Version Date 1 0 15 Mar 2021 Copyright 2021 Thorlabs...

Page 3: ...3 2 6 RS232 Operation 12 3 2 7 UART Operation 12 3 2 8 Reset Button 12 3 2 9 Baud Rate 12 3 2 10 Non Volatile Memory 12 3 2 11 Digital I O Ports 13 3 2 12 Analog Output 13 3 2 13 Autonomous Operation...

Page 4: ...Sensor Connector 22 5 4 Safety 23 5 5 Return of Devices 23 5 6 Manufacturer Address 24 5 7 Certifications and Compliances 25 5 8 Warranty 25 5 9 Copyright and Exclusion of Liability 26 5 10 Thorlabs W...

Page 5: ...king forward to hearing from you Thorlabs GmbH Warning Sections marked by this symbol explain dangers that might result in personal injury or death Always read the associated information carefully bef...

Page 6: ...ace for instrument con trol and to connect a power supply For more information on the special features of the PM103x models please see the table PM103x Model Comparison below For instrument control an...

Page 7: ...ion SMA Connector DA 15 Universal Connector RS232 Operation Analog Output Operation Autonomous Operation to Analog Output Real Analog Output Sensitivity Configurable Analog Output UART Operation 2 Dig...

Page 8: ...nd Connector Housing USB Cable Type A to Mini B with locking screw 1 5m PM103A USB Cable Type A to Mini B with locking screw 1 5m PM103U USB Cable Type A to Mini B with locking screw 1 5m Optional Acc...

Page 9: ...103x can be run with the Thorlabs OPM software Please see the require ments for the OPM on the website Sensor Requirements Thorlabs C Series Sensors The PM103x supports all Thorlabs C Series photodiod...

Page 10: ...the contents for completeness and test the PM103x mechanically and electrically Keep the container for storage or in order to return the product in case of future problems Verify that you have receiv...

Page 11: ...a readout instrument PC using the USB cable or a matching cable for the respective output port SMA or DA 15 For PM103A and PM103 please provide cables Connect a suitable sensor to the PM103x The senso...

Page 12: ...omatically recognize sensors lacking a DE 9 connector as for the case of the A and B series Please contact Thorlabs for an upgrade of old sensors with DE 9 connectors Note Custom Sensors To use sensor...

Page 13: ...g of the micro controller The PM103x is powered either via the USB interface or via the DA 15 interface The status LED indicates appropriate connection to a power supply It turns yellow when the PM103...

Page 14: ...ntrol and data export The PM103 can also be adjusted to accommodate micro controllers using 3 3 V 1 DA 15 Interface 2 Two Digital I O ports 3 NTC input 4 Two Analog Output Ports 5 Real Analog Output 6...

Page 15: ...PM103A PM103A Special Features 1 Autonomous Operation to analog output 2 Configurable Analog Output 3 Configurable Digital In Output 4 Power Meter Control and Output via USB 3 2 4 PM103U PM103U Specia...

Page 16: ...232 operation or to UART operation PM103 uses the DA 15 interface for control and output 3 2 6 RS232 Operation The power meter PM103 can be operated with serial communication via RS232 Please use a RS...

Page 17: ...the USB interface or via the serial interface using SCPI commands or driver functions Please see the document Write Your Own Applications WYOA for PM103x The default Baud Rate is 115200 Note After adj...

Page 18: ...signal is not wavelength or zero corrected and offers the full bandwidth up to 1 MHz without quantization error Note The bandwidth is in accordance with the settings of the sensor and power meter AO2...

Page 19: ...he USB connector or with any supply in the range from 5 V to 36 V via the DA 15 connector Pin 1 and Pin 9 GND As an external power supply we recommend the DS5 5 VDC regulated power supply 3 3 Software...

Page 20: ...e 3 3 2 2 PM103A Specific Software Functions When a PM103A is connected it is possible to configure the analog AO2 SMA port and display the output voltage When using the Scope Mode in the OPM software...

Page 21: ...x is not water resistant Attention To avoid damage to the instrument do not expose it to spray liquids or solvents The unit does not need a regular maintenance by the user It does not contain any modu...

Page 22: ...A Analog Bandwidth DC 150 kHz Dependent on Sensor and Settings Wavelength Correction nm A W Beam Area Setting Diameter 1 e or Rectangular x y AD Converter 16 bit Voltage Input Pyroelectric Sensors Con...

Page 23: ...Corrected Amplified Input Signal Not Corrected NA Voltage Range 0 25 V to 2 5 V 0 25 V to 2 5 V NA Accuracy 3 3 NA Bandwidth up to 1 MHz Depend ent on Sensor and Set tings NA NA Auxiliary Temperature...

Page 24: ...ES100C ES200C Series Photodiodes Max 10 mA Pyro Detectors Max 200 V Status LED Power Sensor PGM Error USB Operating Temperature 0 C to 40 C Storage Temperature 40 to 70 C Dimensions L x W x H 3 80 x...

Page 25: ...nd 12 GND Ground 13 GND Ground 14 N C NTC Thermistor Input Measurement Range 0 1 kW to 100 kW 15 GND RS232 Signal Ground Connect to PC Terminal Ground PC DS 9 Pin5 5 3 Pin Assignment of the Sensor Con...

Page 26: ...e Pyroelectric Sensor NTC 4 PD Photodiode Cathode Input 5 PY Pyroelectric Sensor Input 6 DGND Digital Ground for EEPROM and 5V Output 7 PRESENT Sensor Recognition NTC Input For Custom Sensors Connect...

Page 27: ...radio communica tions However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception wh...

Page 28: ...en closed devices If necessary ask for replacement packaging Refer servicing to qualified per sonnel 5 6 Manufacturer Address Manufacturer Address Europe Thorlabs GmbH M nchner Weg 1 D 85232 Bergkirc...

Page 29: ...5 Appendix Rev 1 0 15 Mar 2021 Page 24 5 7 Certifications and Compliances...

Page 30: ...0Conditions 20of 20Sales_Thor labs GmbH_English pdf 5 9 Copyright and Exclusion of Liability Thorlabs has taken every possible care in preparing this document We however assume no li ability for the c...

Page 31: ...les thorlabs com Thorlabs End of Life Policy WEEE Thorlabs verifies our compliance with the WEEE Waste Electrical and Electronic Equip ment directive of the European Community and the corresponding na...

Page 32: ...www thorlabs com...

Reviews: