5.2 Pin Assignment Output Connector
PM103
: PM103 has a universal DA-15 connector. The Pins are assigned as follows:
Pin
Function
Description
1
Power Supply Pin for Alternative Power Supply with 5 VDC to 36 VDC, when not
powered by USB
2
D I/O 1
General Digital Input / Output port; 3 V logic (Output), 5 V tolerant for In-
put
3
D I/O 2
General Digital Input / Output port; 3 V logic (Output), 5 V tolerant for In-
put
4
AO1
Analog Output; -0.25 V to +2.5 V per Measurement Range
5
AO2
Software Configurable Analog Output; 0 V to 2.5 V
6
N.C.
Not Connected
7
TxD
RS232; Connect to PC terminal RxD (PC DE-9 Pin2)
8
RxD
RS232; Connect to PC terminal TxD (PC DE-9 Pin3)
9
GND
Power Supply Ground
10
GND
Ground
11
GND
Ground
12
GND
Ground
13
GND
Ground
14
N.C.
NTC Thermistor Input; Measurement Range 0.1 k
W
to 100 k
W
15
GND
RS232 Signal Ground; Connect to PC Terminal Ground (PC DS-9 Pin5)
5.3 Pin Assignment of the Sensor Connector
The sensor connector allows the access of all Thorlabs “C-type" photodiode and pyroelectric
sensors. The power meter interface uses the sensor calibration data, stored in the sensor con-
nector, to calculate the corresponding actual power levels.
Additionally, the PM103x is capable to support custom made detectors. Please read the follow-
ing instruction prior to connecting a self made sensor.
Warning
Pin 2 is uniquely used for the EEPROM Digital I/O (memory in Thorlabs sensor heads) and
MUST NOT be used. Connecting this Pin may cause malfunction of the PM103x.
Pin-out of the DE-9 female
connector (female)
Pin
Function
Description
1
Power Supply
+5 V; Max Current 50 mA from this Pin
2
EEPROM
Sensor Memory for the Calibration Data