background image

InGaAs Switchable Gain Detector

 

Chapter 1: Warning Symbol Definitions 

Rev D, April 21, 2020 

Page 1 

Chapter 1  Warning Symbol Definitions 

Below is a list of warning symbols you may encounter in this manual or on your 
device. 

Symbol 

Description 

 

Direct Current 

 

Alternating Current 

 

Both Direct and Alternating Current 

 

Earth Ground Terminal 

 

Protective Conductor Terminal 

 

Frame or Chassis Terminal 

 

Equipotentiality 

 

On (Supply) 

 

Off (Supply) 

 

In Position of a Bi-Stable Push Control 

 

Out Position of a Bi-Stable Push Control 

 

Caution: Risk of Electric Shock 

 

Caution: Hot Surface 

 

Caution: Risk of Danger 

 

Warning: Laser Radiation 

 

Caution: ESD Sensitive Components 

Summary of Contents for PDA20CS2

Page 1: ...PDA20CS2 InGaAs Switchable Gain Detector User Guide...

Page 2: ...4 1 Theory of Operation 4 4 2 Responsivity 4 4 3 Dark Current 4 4 4 Bandwidth and Response 5 4 5 Terminating Resistance 5 4 6 Gain Adjustment 5 Chapter 5 Troubleshooting 6 Chapter 6 Specifications 7...

Page 3: ...Symbol Description Direct Current Alternating Current Both Direct and Alternating Current Earth Ground Terminal Protective Conductor Terminal Frame or Chassis Terminal Equipotentiality On Supply Off...

Page 4: ...rives 50 load impedances up to 5 V The PDA20CS2 housing includes a removable threaded coupler SM1T1 and retainer ring SM1RR that is compatible with any number of Thorlabs 1 threaded accessories This a...

Page 5: ...erformance Connect the remaining end to a measurement device such as an oscilloscope or high speed DAQ card Caution Many high speed oscilloscopes have input impedances of 50 In this case do not instal...

Page 6: ...ed to a photodiode The PDA with Transimpedance Amplifier does control the dark current flowing out Looking at the figure above it can be noted that Point B is held at ground and the amplifier will try...

Page 7: ...t the gain so that the measured signal level out of the PDA20CS2 is below 10 V 5 V with a 50 load to avoid saturation For low terminating resistors 5 k or 1 error an additional factor needs to be cons...

Page 8: ...if using a Voltage measurement device Verify that the optical signal wavelength is within the specified wavelength range Verify that the optical signal is illuminating the detector active area Output...

Page 9: ...Typ 12 mV Max Offset 8 mV Typ 12 mV Max 10 dB Setting 50 dB Setting Gain Hi Z 4 75 x 103 V A 2 Gain Hi Z 4 75 x 105 V A 2 Gain 50 2 38 x 103 V A 2 Gain 50 2 38 x 105 V A 2 Bandwidth3 1 5 MHz Bandwidth...

Page 10: ...Hi Z General On Off Switch Slide Gain Switch 8 Position Rotary Output BNC DC Coupled Package Size 2 79 x 2 07 x 0 89 70 9 mm x 52 5 mm x 22 5 mm PD Surface Depth4 0 09 2 2 mm Weight Detector Only 0 10...

Page 11: ...Switchable Gain Detector Chapter 6 Specifications Rev D April 21 2020 Page 9 6 1 Response Curve 1000 1200 1400 1600 800 1700 0 0 0 2 0 4 0 6 0 8 1 0 1 2 PDA20CS2 Responsivity Responsivity A W Wavelen...

Page 12: ...InGaAs Switchable Gain Detector Chapter 6 Specifications Page 10 TTN135000 D02 6 2 Mechanical Drawing...