![THORLABS MAX200 Manual Download Page 21](http://html1.mh-extra.com/html/thorlabs/max200/max200_manual_1110874021.webp)
20
MAX200 Series Motorized Microscopy Stages
4.2.5 Fitting a 96 Hole Well Plate
This section is applicable only to the MAX202 and MAX203 stages
1) Fit the stage to the microscope as detailed in Section 4.2.2. or Section 4.2.3.
2) The stage is supplied already fitted with cleats to secure the well plate. Slacken
the bolts holding the cleats and turn each cleat head so that the flat edge is facing
the well plate recess - see Fig. 4.4.
3) Fit the well plate into place.
4) Turn the cleat heads through 90° so that the curved edge is inboard, and the well
plate held in place.
5) Tighten the cleat bolts to secure the well plate.
Caution
. The stage is supplied with cleats already fitted, and these should not normally
be removed. When refitting cleats which have previously been removed, use the original
fixings. Do not use fixings that protrude more than 2.5mm into the moving platform as
this could damage the internal mechanism of the stage.
Damaging the stage in this way could invalidate the warranty.
Fig. 4.4 Mounting a 96 Hole Well Plate.
Qty 6 mounting
cleat, turned with flat
edge inwards