8 kHz Galvo-Resonant Scanner and Controller
Chapter 5: Software
Page 16
TTN118194-D02
5.4.
Updating the Firmware
Make sure to update your firmware immediately if you receive a newer version of
LSK-GR firmware from Thorlabs.
1. Save the latest firmware (.hex file) on your computer.
2. Open the
LSK-GR Control
window, and click
File>Update Firmware
.
Make sure you have the controller box plugged into your computer via
USB.
Figure 20 Update Firmware
3. Browse the .hex file on your computer then click
Open
. A
Confirmation
window appears; click
Yes
to continue.
Figure 21 Confirmation Window
4. Click
Start Firmware Update
to install the new firmware.
Figure 22 Start Firmware Update
Note: If bootloader could not be found because Windows was
discovering the device, check
Device Manager
for bootloader and try
Update Firmware
again (step 2).