5) Push the motor into the clamp until the center of the mirror is aligned with the center of the hole in the remaining clamp

when viewed from above.

Fig. 2.4 Aligning the X-axis Motor

6) Repeat items (4) and (5) for the Y-axis motor. Push the motor into the clamp until the center of the mirror is aligned with

the center of the laser exit hole and the X-axis mirror as shown below.  

Fig. 2.5 Aligning the Y-axis Motor

Note

. If alignment of the mirrors is crucial, the system should be mounted within the application at this point, and the mirror

alignment checked experimentally before proceeding.

7) Ensure that the motor connectors are aligned as shown in Fig 1.5, then tighten the pinch bolts on both clamps.

Caution. Ensure that the thermal insulation is fitted. The clamp pinch bolts will not tighten sufficiently without this

material.

  

Fig. 2.6 Tightening The Pinch Bolts

Align the X axis