1) Unpack the cage system adapter and check that all the parts are present.

Fig. 2.1 Cage System Adapter - Unpacked

2) Loosen the pinch bolts on the existing mount, and remove the motor/mirror assembly, taking care not to touch the mirror

surface.

3) Once removed, check that the thermal insulation material is undamaged. Damaged material should be replaced. 

Fig. 2.2 Motor/Mirror Assemblies with Insulation Material

4) Place the cage adapter body on the worksurface as shown opposite, with the open face (base) uppermost. Insert the motor

assembly with the smallest mirror 

(X-axis) into the upper clamp - see Fig 1.3.

Caution. Ensure that the thermal insulation is fitted. The clamp pinch bolts will not tighten sufficiently without this

material.

Gently twist the motor body as it is pushed into the clamp, taking care not to damage the thermal insulation material.

Fig. 2.3 Inserting the X-axis Motor

Adapter body with

motor clamps fitted

Screws Qty 4

Rubber Gasket

Base/cover plate

Reviews: