5 Maintenance and Service
39
2. The fuse holder (see Figure 24) is located below the 3-pole power connector of the mains
jack. Release the fuse holder by pressing its plastic retainers with the aid of a small screw-
driver. The retainers are located on the right and left side of the holder and must be pressed
towards the center.
3. Replace the defective or wrong fuses and press in the fuse holder until locked on both sides.
Fuse type
T0.8A 250V (800 mA, time-lag, 250V)
Note
The fuses must meet IEC specification 60127-2/III, time characteristic: time-lag (T), 250V AC,
size 5 x 20 mm.