THORLABS ELL14 Operating Manual Download Page 3

ELL14 Motorized SM1 Optic Rotator

 

Chapter 1: Introduction 

Rev. B, Feb 2020 

 

Page 1

 

Chapter 1 

Introduction 

The ELL14 is part of the Thorlabs series of resonant piezo motor circuits and bare modules for OEM applications. This SM1-
threaded  Optics Rotator  can be adapted to  30 and 60 mm cage systems  and can also be  adapted for post mounting  –  see 
section 3.1 for more details. 

The resonant piezo design of these motors offers fast response times and precise positioning and is therefore particularly useful 
in scanning applications. 

The high-speed digital signal processing (DSP) architecture supports a multi-drop serial communication protocol, and a set of 
digital IO lines allows the user to control the movement and state manually by switching the lines high (5V) or low (0V).  

The rotator features a central SM1-threaded aperture to accept an optic, and delivers a travel range of 720° rotation. The angular 
position is requested and displayed from 0 to 359.99°. If a rotator is driven past the 359.99° rotation point, the display reverts 
back to zero and counts up to 359.99° again. Furthermore, the unit will not respond to requests for a move to a position greater 
than 359.99° and an error message will be generated. It is designed for closed loop applications requiring rotational positioning 
with 44.0 µrad of resolution. 

Homing is achieved using a combination of a reflecting optical sensor (IR) for coarse (0.5 to 1.0 mm) positioning, then a magnetic 
sensor for fine (1.0 µm resolution) positioning. Using the ELLO software, the user can modify the offset value to shift the homing 
position (up to a ¼ of turn). Furthermore, coarse homing can be selected in a clockwise (CW) or counter clockwise (CCW) 
direction (fine homing is always performed in a CCW direction to guarantee repeatability). 

 

Note 

The module must be powered via an external 5V power supply. A suitable PSU is supplied in the kit. For customers 
wanting to use a third party PSU, the Connector on the interface board is a DC Jack connector, 6.3mm OD (GND), 
2.1mm ID (+5V). 

 

A hand-held controller is supplied with the ELL14K evaluation kit to allow homing and manual jogging and/or positioning. The 
unit can also be driven remotely via PC-based software, downloaded from 

www.thorlabs.com

. A compatible USB driver and 

source code are included in the software download package. 

 

Summary of Contents for ELL14

Page 1: ...ELL14 Motorized SM1 Optic Rotator ELL14 and ELL14K Motorized SM1 Optics Rotator Kit Operating Manual Original Instructions...

Page 2: ...nsation 6 4 2 1 Hand held Controller 8 4 2 2 Software Control 9 4 2 3 Communications Protocol 10 4 2 4 Connecting Multiple Devices 10 4 2 5 Controlling the Rotator Without the Handset 10 Chapter 5 Tro...

Page 3: ...up to 359 99 again Furthermore the unit will not respond to requests for a move to a position greater than 359 99 and an error message will be generated It is designed for closed loop applications req...

Page 4: ...ot specified by the manufacturer the protection provided by the equipment may be impaired In particular excessive moisture may impair operation The equipment is susceptible to damage from electrostati...

Page 5: ...in explosive environments The unit is not designed for continuous operation Typical lifetime is 3 3 million switching operations which equates to 100 km of travel see Chapter 5 for more details Mounti...

Page 6: ...s 3 Possible Mounting Hole Patterns The rotator is shipped with two SM1 retaining rings allowing SM1 optics to be fitted There are several options for mounting the rotator The unit can be mounted dire...

Page 7: ...he rotator can be fitted to our ER series cage rods within a 30 mm cage system Figure 3 ELL14 Mounted in a 30 mm Cage System If combined with the SR series posts and the ELLA1 Adapter the unit can be...

Page 8: ...switch ON A 5 V PSU is supplied with the ELL14K Caution Boot up the PC BEFORE connecting the USB cable DO NOT connect a powered ELL kit to a PC that is not powered up and running 2 Using the USB cabl...

Page 9: ...ming and Jogging functionality can also be accessed by applying voltages to the digital lines on Connector J1 The modes of control are described in the following sections Caution In all modes the angu...

Page 10: ...ternal 5V power supply The external PSU connector allows the rotator to be used in the absence of a PC with control being achieved via the handset buttons LED1 green is lit when power is applied to th...

Page 11: ...r to the rotator unit 3 Connect the hand held controller to the 5V Power Supply and switch on 4 Connect the hand held controller to the PC USB port and wait for the drivers to be installed 5 Run the E...

Page 12: ...e contained in the help file supplied with the software Connect the first device to the PC USB port then run the Elliptec software and load the device Change the address of the first device Connect th...

Page 13: ...IN RX receive 3 3V TTL RS232 4 OUT In Motion open drain active low max 5mA 5 IN JOG Mode active low max 5V 6 IN BW Backward active low max 5V 7 IN FW Forward active low max 5V 8 PWR VCC 5V 10 800mA C...

Page 14: ...ze the buildup of debris on the track and will prevent the motors digging a groove over the most used area of contact Typically a travel cycle should be performed every 10K operations Restoring Factor...

Page 15: ...use an additional 5 V 1 A power supply and a USB connection Integrators should search for optimal frequency on every power up sequence commands s1 s2 see ELLx protocol document How do I restore the f...

Page 16: ...time will depend on several factors e g load number of homing operations number of frequency searches etc and users must take into account all these factors when considering life time For example homi...

Page 17: ...ace board When making a cable to operate multiple devices it is important to observe the correct pin orientation The following procedure offers guidance in making such a cable 1 Gather together the pa...

Page 18: ...ke connection with the ribbon cable 4 If other connectors are required they should be fitted at this point Slide each connector onto the cable paying attention to the orientation as shown below then c...

Page 19: ...to 5 5 V Typical Current Consumption During Movement 800 mA Standby Current 0 05 A Motor Type Elliptec Resonant Piezo 8 Conductor Ribbon Cable Length Supplied 250 mm 8 Conductor Ribbon Cable Length M...

Page 20: ...gainst harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance...

Page 21: ...lie bin logo see right were sold to and are currently owned by a company or institute within the EC and are not dissembled or contaminated Contact Thorlabs for more information Waste treatment is your...

Page 22: ...www thorlabs com...