4 Operation
16
Unless you wish to use a wavefront sensor with the DMH40 to correct a wavefront as de-
scribed in Closed Loop Control, please click okay. Otherwise, please install the WFS soft-
ware and restart the Thorlabs Deformable Mirror software.
8. In case no Deformable Mirror was connected when the software was started, no Deformable
Mirror will be recognized. In that case connect the Deformable Mirror and click "REFRESH".
To connect the recognized device, select it and click "connect".
In case two Deformable Mirrors were connected to the PC, select and connect the Deform-
able Mirror of choice.
Note
Pressing the keyboard button
F1
calls up online help to the Thorlabs Deformable Mirror soft-
ware.
4.5 Out-of-the-Box Test
It is simple to test the DMH40 "out of the box". You do not need even a light source or other op-
tical equipment - just your eyes:
·
Connect the DMH40 to the power supply and to the PC with installed software.
·
Start the Deformable Mirror software. The DMH40 is initialized.
·
Look onto the mirror surface and change the position until you can see your eye's pupil.
·
Press the "Relax" button in the toolbar and observe the diameter of your pupil. The Relax
function changes the mirror segment voltages in such way that the mirror surface altern-
ates between concave and convex, changing the diameter of the reflected pupil of your
eyes.
Summary of Contents for DMH40-F01
Page 1: ...Piezoelectric Deformable Mirror 2021 DMH40 M F01 DMH40 M P01 Operation Manual...
Page 2: ...Version Date 1 1 15 Apr 2021 Copyright 2021 Thorlabs...
Page 56: ...2021 Thorlabs 51 DMH40 7 4 Dimensions DMH40 xxx Imperial...
Page 59: ...2021 Thorlabs 7 Appendix 54 7 8 Certifications and Compliances...
Page 66: ...www thorlabs com...