Rev A Jan 2020
Page 7
Chapter 4 Operation
2) 150 V peak-to-peak, 703 Hz sine wave into 10.4 µF capacitive load.
150 V peak to peak sine wave = 53 Volt RMS
At 703 Hz, the impedance of the 10.4 µF capacitor is -21.8
(theoretical)
Calculated RMS current: 2.43 A
3) 150 V peak to peak, 4 kHz square wave into 1.1 µF capacitive load.