Chapter 3 Getting Started
Chapter 3 Getting Started
3.1 Siting
The unit is designed to be used free standing on a shelf, benchtop or similar surface.
3.2 Environmental Conditions
Location
Indoor use only
Maximum altitude
2000 m
Temperature range
5 °C to 40 °C
Maximum Humidity
Less than 80% RH (non-condensing) at 31°C
Line voltage fluctuations
Less than ±10% of the line voltage
To ensure reliable operation the unit should not be exposed to corrosive agents or excessive moisture, heat or dust.
If the unit has been stored at a low temperature or in an environment of high humidity, it must be allowed to reach ambient
conditions before being powered up.
3.3 Electrical Installation
3.3.1 Connecting To The Supply
Caution
Ensure that proper airflow is maintained to the side and rear of the unit. Insufficient air flow will cause overheating and
premature failure.
Do not cover the ventilation slots/fans on the side and rear of the unit. Place the amplifier in a location with adequate
ventilation to prevent internal heat build-up. The device needs to be installed horizontally with 3 cm air circulation area.
Vertical mounting prevents internal convection.
When siting the unit, it should be positioned so as not to impede the operation of the rear panel power supply switch.
Warning
Operation outside the following environmental limits may adversely affect operator safety.
Caution
In applications requiring the highest level of accuracy and repeatability, it is recommended that the amplifier unit is
powered up approximately 30 minutes before use, in order to allow the internal temperature to stabilize.
Warning: Risk of Electrical Shock
The unit must be connected only to an earthed fused supply of 85 V to 264 V.
The unit is shipped with appropriate power cables for use in the UK, Europe and the USA. When shipped to other
territories the appropriate power plug must be fitted by the user. Cable identification is as follows:
Brown: Live
Blue: Neutral
Green/Yellow: Earth/Ground
Caution
High electrical currents require suitable connectors and cabling. The unit is fitted with SHV connectors which offer a high
degree of safety and protection. A suitable cable (CA3262) is available separately from www.thorlabs.com. A suitable
mating connector (T4004, T4005 or T4006) is available separately for use in connecting the piezo device being driven.
It is not recommended to use any other type of connector or adapter.
In any case, the level of insulation provided by the connection must be double/reinforced insulation at 150 V peak
minimum.