5.3 Replacing the mains fuse
LDC2xx / page 26
5.3 Replacing the mains fuse
If the mains fuse has opened due to line distor tions, incorrect line voltage or other
causes, it can be easily replaced from the rear without opening the unit.
Attention
To avoid risk of fire only the appropriate fuse for the corresponding
line voltage must be used.
1. Turn off the LDC2xx and disconnect the mains cable.
2. Open the fuse drawer in the main
s connector R9 (see Figure 7) with a
screwdriver.
3. Replace the defective fuse (one spare fuse is included in the fuse holder) and
close the drawer.
LDC200V and LDC201U:
100 V
250 mA, Slow, 250V
T0.25A250V
115 V
250 mA, Slow, 250V
T0.25A250V
230 V
160 mA, Slow, 250V
T0.16A250V
LDC202B, LDC205B, LDC210B and LDC220:
100 V
500 mA, Slow, 250V
T0.5A250V
115 V
500 mA, Slow, 250V
T0.5A250V
230 V
250 mA, Slow, 250V
T0.25A250V
All fuses are to be IEC 60127-2/III
Figure 7
Changing the mains fuse