3.3 Selecting the laser diode polarity
LDC2xx / page 16
Switch off the LDC2xx.
Select the desired polarity with the switch "LD POL" (R6).
For laser diodes with a built-in monitor diode the common pin of laser diode and
photodiode is the ground pin.
The selected polarity of the output is indicated with the LED "AG" (L1) or "CG" (L2).
For safety reasons there is a mid position of the switch "LD POL" (R6).
NOTE:
If the mid position of the switch "LD POL" (R6) is selected, neither the LED
(L1) nor (L2) lights up. The laser current cannot be switched on.
If the laser diode is connected correctly (refer to chapter 3.1.3) but the polarity of the
laser diode selected wrong with switch (R
6) the laser diode current cannot be
switched on.
If the position of the switch "LD POL" (R 6) is changed with the laser current switched
on the laser diode current is switched off automatically.
NOTE:
For safety reasons you should select the polarity for the laser diode only
with the laser diode current switched off.