THOMSON DFT-6900 User Manual Download Page 6

DFT-6900 Battery Activator Instruction Manual

 

 

1.6 Main Parts 

The  main  unit  comes  with  two  current  clamps  (one  red  and  one  black)  and  one  voltage  detection  line.  Connect  the 

three-pin terminal of the voltage detection line to the “voltage detection” terminal on the back panel of the main unit. 

The terminals of the red current clamp are connected to the red terminal block on the back panel, and the terminals of 

the  black  current  clamp  are  connected  to  the  black  terminal  block  on  the  back  panel.  Note:  It  is  strictly  forbidden  to 

reverse connection! 

 

1.7 Operation Instruction 

Touch screen for all operation 

1.8 Wiring 

Activation,  charge/discharge  wiring: 

Connect  the  current/voltage  clamps  with  the  battery  positive/negative 

poles:  Black  clamp  to  negative  pole,  red  clamp  to  the  positive  pole.  (Note:  the  positive  and  negative  poles  are 

strictly prohibited to be reversed connection). If the current clamp is reversed, it will cause serious damage to the 

instrument. If the voltage clamp is reversed connection, the instrument cannot be operated normally, at the same 

time,  the  instrument  will  prompt  the  user  to  convert  the  (voltage)  clamps,  but  it  will  not  cause  damage  to  the 

instrument.

 

 

Current clamp reversed connection not allowed to battery poles; otherwise the instrument 

will be seriously damaged. 

Summary of Contents for DFT-6900

Page 1: ...DFT 6101 Battery Impedance Tester Instruction Manual DFT 6900 Battery Activator Users Manual THOMSON TECH CO LTD www thomsontesting com...

Page 2: ...on Instruction 5 1 8 Wiring 5 Chapter 2 Functions operations 6 2 1 Discharge 6 2 2 Charge 7 2 3 Activator 8 2 3 1 Activator Setting Interface 8 2 3 2 Activating Process 8 2 4 Data Management 8 2 4 1 D...

Page 3: ...ct operation of the instrument 1 To avoid electrical shock or fire read these precautions first before using the unit 2 Except as explained in this manual do not attempt to service this equipment your...

Page 4: ...ing and activation The curve can be displayed after test The right side of the screen shows the charge discharge curve and the left side shows the charge discharge or activation parameters including v...

Page 5: ...via a U disk By transmitting the data to the PC the battery quality can be monitored for a long time Through the analysis of the battery operating status data various problems which affect the battery...

Page 6: ...ion 1 7 Operation Instruction Touch screen for all operation 1 8 Wiring Activation charge discharge wiring Connect the current voltage clamps with the battery positive negative poles Black clamp to ne...

Page 7: ...e Click Discharge to enter parameters setting It is necessary to set six parameters battery number battery type rated capacity cutoff voltage discharge current and discharge duration Set the discharge...

Page 8: ...rged capacity the charged time and the stop parameters Note 1 The cut off voltage set in the parameter setting is to prevent the battery from over charging discharging which leads to battery damaged C...

Page 9: ...battery The activation settings interface is as follows After clicking the up and down direction keys to set the Cycle No you can set the parameters of each activation by clicking the page up and page...

Page 10: ...e U disk first and the green U symbol will be displayed in the upper right corner of the menu Note U disk read and writes error 1 U disk plugged in properly 2 It is strongly recommended that users use...

Page 11: ...2 4 4 Data Formatting In the data management menu click the Format option to delete all data 2 5 System setting In the main menu interface click System Settings to enter the system management interfa...

Page 12: ...he U disk to Tester After the software update is completed the Tester enters the boot interface and can perform normal operation 2 5 3 Language Enter the System Settings menu page click the Language S...

Page 13: ...ents or alcohol as they can deform or discolor the Tester and the accessories 3 1 2 Clamps Maintenance Clean Clean the main unit and its accessories with damp cloth and a mild soap 3 2 Storage After u...

Page 14: ...e N cycles to prolong the time Example If the charging required time is 28 hours set the activation cycle 2 the first charging time is 18 hours the discharge duration is 0 the second charging time is...

Page 15: ...disk cannot be detected after it is inserted to the PC The U disk may not be well inserted If so reinsert it correctly The U disk may not be in FAT format If so format the U disk And if other problem...