SCPI Command Set
Syntax
:INPut:GAIN:IF <NR1 [unit]>
:INPut:GAIN:IF? [MAX | MIN]
Parameter
<Gain value [unit]>
Input Data Type
Signed integer [character]
Allowable Values
0 to 30
Query Response
<NR1>
Output Data Type
Integer
Default I/O unit
dB
*RST State
0
Examples
:INPUT:GAIN:IF -2
:INP:GAIN:IF 20 dB
:INP:GAIN:IF?
:INPut:GAIN:HDR
This command sets or queries variable NB IF gain of the HDR signal path.
Syntax
:INPut:GAIN:HDR <NR1 [unit]>
:INPut:GAIN:HDR? [MAX | MIN]
Parameter
<NR1 [unit]>
Input Data Type
Signed integer [character]
Allowable Values
-10 to 34
Query Response
<NR1>
Output Data Type
Signed integer
Default I/O unit
dB
*RST State
25
Examples
:INPUT:GAIN:HDR -5
:INP:GAIN:HDR 20 dB
:INP:GAIN:HDR?
:INPut:MODE
This command sets or queries the WSA5000's RFE mode of operation.
Notes:
The RFE modes affect the data packing method due to the different output data
width. The type of DSP applied would also change the data output of some of the modes
as well. For example, SH mode with frequency shift would change from I
14
only data to
I
14
Q
14
data output. The VRT's Stream ID would identify the format accordingly. See
2: Radio RFE Modes and DSP Data Output Formats
).
for the IBW of each mode and the related notes.
CONNector is to be used with SH or SHN mode, see the
Important Note listed under that command.
62
ThinkRF WSA5000 Wireless Signal Analyzer Programmer's Guide