SCPI Command Set
Syntax
[:SENSe]:LOCK:REFerence?
Parameter
None
Query Response
0 | 1
1
Reference PLL is locked
0
Reference PLL is not locked
Output Data Type
Integer
*RST State
N/A
Example
LOCK:REF?
[:SENSe]:LOCK:RF?
This command queries the lock status of the RF VCO (Voltage Control Oscillator) in the
RFE.
Syntax
[:SENSe]:LOCK:RF?
Parameter
None
Query/Response
0 | 1
1
RF PLL is locked
0
RF PLL is not locked
Output Data Type
Integer
*RST State
N/A
Example
LOCK:RF?
OUTput Commands
:OUTput:IQ:MODE
This command sets or queries the WSA5000's IQ output path to use the digitizer section
for data output or the direct output to the IQ connector port of the WSA.
Important Notes:
- When the CONNector option is selected:
+ all commands affecting the digitizer data path of the WSA5000 will not apply, these
include
) and digital signal
processing (
).
+ no VRT context packets will be sent out from the WSA.
SH or SHN, the spectral inversion solution as
is required depending on the frequency input. The spectral
inversion indicator is available through the GPIO port. Contact ThinkRF's Support for
further details.
- The CONNector mode only works on certain WSA5000 model. Verify that your product
supports this method. A system error will be returned if the model does not accept this
option.
68