Remark:
Events can, possibly, be activated with the cleaning work; they should be considered then
with the evaluation / further processing.
7 Settings
7.1 Setting of Incidences and Switch-off Delay
The criterions of signal output for start
and end of the precipitation are factory-
set.
If the factory-setting has to be changed due
to special ambient conditions this can be
done by means of switches
DIP1 and
DIP2
.
DIP1 = Switch-off delay (for the end
of precipitation period).
DIP2 = Drop event filter (for the start
of precipitation period).
Settings see
table 1 or 2
Process:
-
Interrupt power supply and signal
current.
-
Remove 5 screws at the bottom
side of the instrument.
-
Remove upper part (cover).
-
Set switch DIP1 / DIP2 acc. to table
1 or 2.
-
Put on the upper part again and
screw it evenly and tightly to the
bottom.
Attention:
The screws for the cover must be tightened
by a torque of
1Nm to 2Nm
.
Make sure, that the contact pins are not
deformed when putting on the cover.
-
Restart instrument.
Figure 1
: Position of DIP – switch
DIP1
DIP2