P|11
Displayed
Code
Meaning
Suggested Solution
Low Battery
Replace batteries as soon as
practical.
Very Low
Battery
Replace batteries prior to further
use.
Ambient
Temperature
too LOW
Allow Caregiver® to warm to
temperature above
50
0
F (10
0
C).
Ambient
Temperature
too HIGH
Allow Caregiver® to cool to temperature
below 104
0
F (40
0
C).
Sensor Error
or Ambient
Temperature
Unstable
Allow Caregiver® to equilibrate to
the current room temperature for 20
minutes. Contact Thermomedics®
Support if this error persists.
Body Mode
Temperature
is below 94
0
F
(34.4
0
C)
Remove Sensor Lens cap. Allow
patient to warm up in ambient
room temperature above 50
0
F
(10
0
C).
Body Mode
Temperature
is above 108
0
F
(42.2
0
C)
Repeat the temperature. If this
condition persists, the ambient
temperature of the thermometer
may be outside the acceptable
limits. Allow to cool for
20 minutes.
Displayed
Temperatures
Appear to be
Too Low
Remove Sensor Lens Cap. Allow
thermometer and patient to
adjust to room temperature for 20
minutes. Remove perspiration, oil
and makeup from forehead. Also
consult the knowedgebase at
www.thermomedics. com
Troubleshooting - Error Codes