Copyright 2018 Thermoline Scientifi c Equipment Pty Ltd
Remove the equipment from the packing material and check for damage. Notify the detail of any damage to your supplier or
to Thermoline Scientifi c without delay.
Retain the packing materials until the equipment has been thoroughly tested.
Inventory:
• Shelves and shelf support clips include std quantities.
• 1 x 3-pin plug for connection of alarms to remote Building Monitoring System (BMS).
• Door keys.
• Chart recorder keys and paper. (Only if optional chart recorder is supplied).
The Thermoline range of Laboratory refrigerators are designed to store temperature sensitive products within an adjustable
temperature range. This product is identifi ed by a label on the front of the unit as shown below:
This products’ intended use is for the storage of temperature sensitive products. It CANNOT be used as a pharmacy or
vaccine refrigerator.
All TELR/TLR Refrigerators have loading requirements. When loading, product should be evenly distributed throughout the
refrigerator. It is important to leave 1/3 of each shelf level free (3cm front, side, rear and between products) this allows air to
fl ow evenly throughout the cabinet and maintain temperature accuracy in the refrigerators workspace.
The digital temperature control monitors product temperature, and includes a temperature logging feature that allows the
operator to determine the maximum and minimum temperatures reached during daily operation.
When correctly maintained, the refrigerator has an expected lifespan in excess of 5 years. Regular maintenance and checks by
a licensed refrigeration technician will help extend the life of the unit.
The display also incorporates the HI/LO temperature alarms emitting both visual and audible signals, which can also activate a
building monitoring system (BMS), via a connection socket on the cabinet.
High temperature alarm set at 1ºC above setpoint.
Low temperature alarm set at 1º C below setpoint.
Operating temperature set b4 & +10º C.
These cabinets will operate from a standard 10 amp, 240 volt, 50Hz, general purpose outlet (power point). Cooling is by means
of a hermitically sealed refrigeration unit using environmentally friendly refrigerant.
Fan forced air circulation internally ensures that even temperature conditions are obtained on all shelf levels. The controller
monitors the temperature inside of the workspace. The control sensor is located in a bottle that simulates product, this give a
more accurate reading of the temperature inside of the workspace.
UNPACKING
USE & FUNCTION
4
Laboratory Refrigerator