3. Connect the capillary to the flow cell inlet.
Figure 21: Capillaries properly connected in the detector
5.6.4.2
Connecting the Detector Waste Line
Preparations
1. See
section 5.6.4 Flow Connections to the Flow Cell, page 67
2. On the system base, route the detector waste line through the
dedicated detector waste outlet in the base to waste. Follow the
instructions in the
Vanquish System Operating Manual
.
Follow these steps
1. On the flow cell outlet, connect the waste line. Make sure that you
secure the waste line appropriately in the capillary clips on the
detector front panel.
2. On the partition panel, route the waste line through the top recess.
Make sure that you use the suitable partition panel plug.
3. From the Vanquish system base, route the waste line upward to the
detector, through the tubing guides of the system modules below
the detector.
Make sure that the waste line has been connected to the waste
container and routed through the system base as described in the
Vanquish System Operating Manual
.
Summary of Contents for Vanquish VH-D10
Page 4: ...Contacting Us Page 4 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 44: ...4 Unpacking Page 44 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 80: ...5 Installation Page 80 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 110: ...6 Operation Page 110 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 162: ...7 Maintenance and Service Page 162 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 190: ...11 Appendix Page 190 Vanquish Diode Array Detectors VH D10 Operating Manual...
Page 191: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 191 12 Index...
Page 195: ...12 Index Vanquish Diode Array Detectors VH D10 Operating Manual Page 195...