Troubleshooting
Diode Array/Multiple-Wavelength Detectors (VC-D11, VC-D12, VF-D11)
Operating Manual
Page 159
Messages in the Chromeleon Audit Trail are preceded by an icon. The
icon identifies the seriousness of the problem (refer to the
Chromeleon
Help
). For possible causes and remedial actions, see
Firmware Failure
If a firmware failure occurred during operation of the module, an
exception log has been created about the processes during the firmware
failure. The firmware sends the exception log to the Audit Trail when the
module is connected in the Chromeleon software.
In this case, observe the following:
•
Send the Audit Trail as
.cmbx
file to the Technical Support before
you clear the log.
•
To clear the exception log and continue operation of the module,
perform the command
ExceptionLogClear
.
For more information, refer to the
Chromeleon Help
.