4  |   Installation

SpeedVac™ SPD130DLX

Installation

Contents

Unpacking

. Open the shipping cartons. Carefully remove the 

instrument and accessories. 

Lift and carry with two 

people, holding securely underneath with both hands. 
Use proper lifting technique (lift with the legs, not the 
back) to avoid personal injury

. Compare the contents with 

the packing list. If there is a discrepancy, call Thermo 
Scientific technical service. 

Inspection

. Inspect the unit and accessories for damage that 

may have occurred during shipment. Should there be any 
damage, report it to the carrier and contact Thermo Scientific 
immediately. Make sure the carrier inspects the damage and 
leaves an inspection report. Register any claims for shipping 
damage against the carrier or his agent. Save the shipping 
carton in the event a return is necessary. Call Thermo 
Scientific technical service for further assistance. 

Site preparation

. The SPD130DLX requires a stable surface 

that is clean, dry, level, and within 4 feet (1.2 meters) of a 
compatible electrical outlet. Place the unit in a convenient 
location with access to a vacuum source and a vapor trap

1

.

The SPD130DLX units configured for 115 VAC, 60 Hz, should 
be plugged into a circuit rated for at least 6 amps. The 
SPD130DLX units configured for 230 VAC, 50 Hz, should be 
plugged into a circuit rated for at least 3 amps.

Note: 

Do not use any detachable power cord that is not 

adequately rated for the unit.

Installation of SPD130DLX

Figure 1. Rear View of System

Table 1. Contents

Qty. 

Description

1

Thermo Scientific Savant SpeedVac™ 
SPD130DLX Concentrator

1

Teflon

®

 tubing 1/2” I.D. x 3’ long

1

Line Cord

1.To prevent damage to pump due to vapor exposures from sample solvent and to extend pump life, a vapor trap that condenses and traps vapors must be 
placed in line between the SpeedVac™ concentrator and vacuum pump. We offer integrated universal vacuum systems (integrated vapor trap with the 
vacuum pump) or component vacuum system (vapor trap plus vacuum pump).

CAUTION: 

Be sure to leave at least 4 inch 

clearance on all sides of the unit. Overhead 
clearance, equal to 18 inch (46cm), is required for 
raising the cover. Verify that the unit is on a 
leveled and stable platform. If necessary, move 
the unit to a more suitable location.

WARNING: 

Before connecting the unit to an 

electrical outlet, make certain that voltage, 
frequency, and amperage match the 
requirements indicated on the product label, 
name plate of the instrument. Use sockets with a 
protective earth conductor and correct mains 
cable.

CAUTION: 

Cold trap should be in line between 

the vacuum source and the SPD Unit.

 

Clamps

RVT5105

Refrigerated Vapor
Trap

SPD130DLX

SpeedVac
Concentrator

Vacuum Pump

Summary of Contents for SpeedVac SPD130DLX

Page 1: ...Installation and Operation 80302133 Revision A July 2018 SpeedVac SPD130DLX Vacuum Concentrator...

Page 2: ...mo Fisher Scientific Inc provides this document to its customers with a product purchase to use in the product operation This document is copyright protected and any reproduction of the whole or any p...

Page 3: ...4 Installation of SPD130DLX 4 Operation 5 Description Of Control Panel 5 Manual Run 7 Auto Run Time 8 Auto Run Vacuum 8 Drying Rate 8 Preset Program Settings 9 Exporting Live Run Data 9 HyperTerminal...

Page 4: ...th dual timer for automatic or manual control of run conditions it also features preset and user defined programs which enable user to retrieve and store application protocols for future use The compa...

Page 5: ...dicates extreme low temperatures and high risk of frostbite Do not touch bare metal or samples with unprotected body parts This symbol indicates possible pinch points which may cause personal injury T...

Page 6: ...tive Power 115 VAC 60 Hz 6 amps 230 VAC 50 Hz 3 amps Vacuum Chamber TEFLON coated chemical resistant aluminum casting Cover Heated cover with radiant lamps includes a safety interlock Induction Motor...

Page 7: ...130DLX units configured for 230 VAC 50 Hz should be plugged into a circuit rated for at least 3 amps Note Do not use any detachable power cord that is not adequately rated for the unit Installation of...

Page 8: ...hand to ensure alignment of the pin with the groove Secure the assembly by screwing the retaining knob into the drive shaft above the rotor Tighten it firmly but not excessively Description Of Contro...

Page 9: ...meters is shown by the corresponding indicator that is illuminated 8 Minus Decrements the value of the selected parameter 9 Plus Increments the value of the selected parameter 10 View During a run pre...

Page 10: ...tor with the supplied knob hand tight Close cover 10 Pre heat may be selected at this time to apply a 45 C heat to the chamber 11 Press the Manual Run button The cover locks and the lid locker indicat...

Page 11: ...and remove samples Auto Run Vacuum 1 Refer to the Manual Run section for start up 2 To execute an Automatic Vacuum run execute the following steps a Select the Run Type as Vac b Use the Select button...

Page 12: ...C The heat time is expressed as a 3 digit integer in minutes or CCC for continuous heating The runtime is expressed as a 3 digit integer in minutes The vacuum pressure is represented as a floating poi...

Page 13: ...id stay mechanism that will assist the user in opening and closing the cover of the unit It will secure the lid and maintain it open allowing the user to operate the unit with both hands Figure 3 Lid...

Page 14: ...cid Ammonium hydroxide n butyl alcohol Chloroform Cyclohexane DCM Ethyl acetate Ethanol Ethyl ether n Hexane n Heptane Isopropyl alcohol Isobutyl alcohol Methanol Petroleum ether n propyl alcohol Pyri...

Page 15: ...0mm RH32 13 15 8 17 5 X 102 mm 15ml Corex tubes RH8 17 5 10 8 18 X 100 mm 17 x 95 16 x 100 RH8 18 CENTRIFUGE TUBES 12 10 15ml conical 16 x 120 mm RH10 15 40 6 50ml conical 28 x 115 mm RH6 50 FLASKS 80...

Page 16: ...neutralizing solution for Oligo Preps Used on UVS850DDA only SCT120 Chemical trap order cartridges separately DTK120R Chemical trap kit for radioactivity compatible with all Thermo vacuum concentrator...

Page 17: ...Pump ANT100 Exhaust recovery Post trap Vacuum Tubing From Cold Trap to Back of SPD130DLX UVS850DDA Vacuum Tubing From Back of SPD130DLX To Cold Trap SpeedVac Concentrator SPD130DLX Vacuum Tubing RVT45...

Page 18: ...MA number issued by Thermo Costs of shipping the unit are not covered under warranty The warranty obliges you to follow the precautions in this manual When returning apparatus that may contain hazardo...

Page 19: ...riciclato tramite queste medesime Ulteriori informazioni sulla conformit di Thermo Fisher Scientific con queste Direttive l elenco delle ditte di riciclaggio nel Vostro paese e informazioni sui prodo...

Page 20: ...30 France 33 2 2803 2180 33 2 2803 2180 Germany 0800 1 536 376 49 6184 90 6000 0800 1 536 376 Italy 32 02 95059 552 39 02 95059 552 432 254 375 Netherlands 31 76 579 55 55 31 76 571 4440 Nordic Baltic...

Reviews: