OdaLog
®
Gas Logger User Manual
P/N: 10-1000
Revision: F, Issued: October 2019
Page: 21
Place Dry Pack (moisture absorption pack) [Part No. 12-0129] on top of
-ve
battery clip
Plug top cap back into circuit board and insert top cap (carefully aligning the Top Cap and the
body
If prompted reset OdaLog
®
Date and Time (Refer to the “Setting the OdaLog’s Date and Time”
section in the
OdaStat-G™
software package)
**Note: Some sensors will require up to 24hrs to ‘settle’ after battery replacement.**
Figure 4 - Top Cap Alignment
8.1.2 Filter Replacement
Figure 5 - Filter Replacement
Ensure OdaLog
®
is off
Using the Allen Tool [Part No. 11-0000] remove the three retaining screws from the filter
housing assembly (See Figure 5)
Carefully remove filter housing assembly
Remove and retain o-ring
CAUTION
Ensure the following procedure is only carried out in a non-hazardous, dirt and
grease free area.
NOTE:
Ensure the Top Cap is aligned correctly - See Figure 4