6.10 Switching off
Press & hold button down and wait for display to cycle through screens until “OFF2” is displayed. Keep button
depressed until the display counts down from “OFF2” to “OFF” then release button.
When the button is correctly released to turn the instrument off, the OFF title will shift 1 character to the right to
confirm that the instrument is now in the off state.
Note 1:
This can only be done when logging is stopped unless Auto Log Start is active. (See Section 6.7)
Note 2:
Releasing the button during the countdown routine will cause exit to instantaneous screen.
6.11 Clear Logged Data
Switch instrument “OFF”. Press & hold button whilst “On” counts down from 3 and
continue holding
until
“CLEAR LOG” appears on screen, then release button.
“Y” and “N” will appear on the display. With the magnetic base of the Allen Tool provided select “Y” to clear
the log memory, “DONE” should appear briefly on the screen before returning to the “OFF” state.
Selecting “N” will return to the “OFF” state without clearing the log memory.
WARNING:
All logged data will be lost when the log is cleared, be sure to download any
data that you wish to keep
before
clearing the log.