![ThermoFisher Scientific HERASAFE 2030i Series Operating Instructions Manual Download Page 54](http://html1.mh-extra.com/html/thermofisher-scientific/herasafe-2030i-series/herasafe-2030i-series_operating-instructions-manual_1105593054.webp)
| Operation
44
|
50155215
Herasafe 2030i / Maxisafe 2030i
Chapter 6
Info & Health Status Area
The Info & Health Status Area at the top of the display window appears in all screen
displays. Figure 26 shows an example of the Info & Health Status Area on a unit that is in
a proper working condition.
Figure 26
Info & Health Status Area
The Info & Health Status Area displays the following information:
Unit Name (top left in Figure 26), see “Setting the Unit Name” on page 83.
Date/Time (bottom left in Figure 26), as entered during the initial setup (see “Initial
Setup” on page 37) or adjusted later on (see “Setting the Date” on page 79 and
“Setting the Time of Day” on page 80).
The “Unit Health Status” icon (center in Figure 26) is a touch-sensitive area that opens a
Health Status pop-up window on top of the Main Screen. Four different icons may appear
in this place, representing the overall health status of the biological safety cabinet:
The “Login” prompt (top right in Figure 26) appears only if the unit is set up to
operate in Secure Mode (see “Access Control” on page 112).
Main Screen (Region R2)
The Main Screen takes up the main part of the “Display Settings” screen and displays the
status details on the biological safety cabinet.
Thermo Fisher BSC
11:42 AM | 08/10/2018
Login
The green “Health Status Good” icon indicates that the unit is in
proper operating condition. For more details, see “Viewing Gen-
eral Status Information” on page 70.
The yellow alert triangle is a “Health Status Alert” icon. It indi-
cates that there is an issue that does not stop operation of the
biological safety cabinet immediately, yet requires the operator‘s
attention soon, such as a HEPA filter to be replaced. For more
details, see “Handling Alerts” on page 70.
The “red bell” “Health Status Alarm” icon indicates that there
are one or more major alarm conditions that need to be cor-
rected because they may constitute a hazard to the operator or
the unit itself. The blue circle indicates how many alarms exist.
A red bell with sound waves on both sides (top) means that the
alarm is sounding, while the red bell with the diagonal line
across (bottom) shows that the alarm has been snoozed.
For more details, see “Silencing and Handling Alarms” on page 67.
Summary of Contents for HERASAFE 2030i Series
Page 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Page 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Page 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Page 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Page 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Page 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Page 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Page 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Page 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Page 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Page 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Page 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...