Chapter 3
Operating Valves Automatically
Operation
Operating Manual
41
If you want the valve to switch back before the end of run, press
CLEAR
to return
to the
SELECT PARAM to add
menu and select the same valve again. Press
ENTER
. Type the desired switching time and press
ENTER
. Select the position the
valve is to return to by selecting
ON
or
OFF
and press
ENTER
. With no run table
entries selected, press
CLEAR
multiple times to exit to the run event table.
OPERATING PROCEDURE
Heating the Valve Oven
1. On the TRACE GC Ultra keypad, press
AUX
.
2. Scroll to
Temperature Zones
and press
ENTER
.
current value
target value
3. Enter the target temperature and press
ENTER
.
NOTE
Press
INFO
to determine temperature limits.
AUXILIARY ZONE
Valve oven
43
175
Summary of Contents for Valve Oven
Page 1: ...Valve Oven Operating Manual ...
Page 18: ...Chapter 1 Valve Oven Overview Valve Oven Components 18 Operating Manual ...
Page 34: ...Chapter 2 Valve Installation Configuring Valves 34 Operating Manual ...
Page 42: ...Chapter 3 Operation Operating Valves Automatically 42 Operating Manual ...
Page 46: ...Chapter 4 Troubleshooting 46 Operating Manual ...
Page 52: ......
Page 60: ...Index 60 Operating Manual ...