Chapter 3
Operation
Operating Valves Automatically
40
Operating Manual
6. Press
CLEAR
three times to exit to the
Run Event
table.
NOTE
Pressing
CLEAR
within a run table entry will delete that entry. To use
CLEAR
to back out
of the menu, you must first be out of the run table entries themselves.
Adding a Gas Switching Valve Event to the Run Table
For switching valves, two run time entries are required—one to switch the valves
on and one to switch the valves off.
1. On the TRACE GC Ultra keypad, press
RUN TABLE
.
2. Scroll to
Add run time event
and press
ENTER
.
3. Select
Valve
and press
ENTER
.
4. Scroll to the Switching valve that you want to add and press
ENTER
.
The
Run time
parameter is the time into the run when the valve switches to the
new position. The
Setpoint
parameter is the position that the valve will go to at
the selected time. The
On
value usually designates a clockwise switch and the
Off
value designates a counterclockwise switch.
If you want the valve to switch back to its original position at the end of run, then
do not need enter a second run table entry.
SELECT PARAM to Add
Valve #1
Sampling
Valve #2
Switching
<
Valve #3
Sampling
RUN TIME EVENT
Valve #2
Switching
Run time
0.00
Setpoint
On (Off)
Summary of Contents for Valve Oven
Page 1: ...Valve Oven Operating Manual ...
Page 18: ...Chapter 1 Valve Oven Overview Valve Oven Components 18 Operating Manual ...
Page 34: ...Chapter 2 Valve Installation Configuring Valves 34 Operating Manual ...
Page 42: ...Chapter 3 Operation Operating Valves Automatically 42 Operating Manual ...
Page 46: ...Chapter 4 Troubleshooting 46 Operating Manual ...
Page 52: ......
Page 60: ...Index 60 Operating Manual ...