![Thermo EPD Mk2 Technical Handbook Download Page 30](http://html.mh-extra.com/html/thermo/epd-mk2/epd-mk2_technical-handbook_1104222030.webp)
EPD/HB/40521/000
ISSUE 5
EPD TECHNICAL HANDBOOK
Page 20
3.2.2
Dose Alarms
Dose alarms are checked and updated every second, except at low dose rates
where this period increases to a maximum of 14 seconds. Alarms can be adjusted
or inhibited via the IR communications link.
Dose alarms are calculated against a preset alarm threshold set in the EPD. When
the dose equals or exceeds the dose threshold the LED will illuminate, the sounder
will activate and the appropriate alarm flag on the LCD will be displayed. Note that it
is dose and not total dose that is compared with the threshold.
The alarm may be muted by pressing and holding the button, unless alarm mute (user
silence) is inhibited.
The four most significant digits of the alarm thresholds can be viewed on the EPD
LCD and chang ed using the function button (see Section 3.5.8).
3.2.3
Dose Rate Alarms
Dose rate alarms are checked and updated every second, except at low dose rates
where this period increases to a maximum of 14 seconds. Alarms can be adjusted
or inhibited via the IR communications link.
The dose rate alarms can be made to work with hysteresis, i.e. the alarm 'off'
threshold is lower than the alarm 'on' threshold. The alarm 'off' threshold can be set in
the EPD over the same range as can the alarm 'on' threshold. As the dose rate rises
above the alarm 'on' threshold the LED will flash and the sounder is activated (see
Figure 3.3). The alarm may be muted by pressing and holding the button, unless
alarm mute (user silence) is disabled. The alarm will self-cancel when the dose rate
falls below the alarm 'off' threshold.
Summary of Contents for EPD Mk2
Page 6: ......
Page 10: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page xii ...
Page 14: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 4 This page is intentionally blank ...
Page 25: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 15 ...
Page 51: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 41 This page intentionally blank ...
Page 52: ......
Page 57: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 47 Figure 4 1 EPD Dimensions ...
Page 86: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 76 This page is intentionally blank ...
Page 103: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 93 This page is intentionally blank ...
Page 104: ......
Page 110: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 100 This page intentionally blank ...
Page 112: ...EPD FAULT LOG FORM HB40521 ...