ARL 3460 Metals Analyzer
Page 9/17
Technical Description
2
1.8
Storage
If laboratory is not ready when the instrument arrives, the instrument must be stored in an
adequate room. Recommendations are as follows:
The storage room must be closed, dry and must have at least a cemented floor.
Temperature:
min. +2°C, max. 45°C
Humidity:
max. 80%, without condensation
The storage room must be free of dust and corrosive vapours.
The instrument must stay on its transportation pallet. This makes subsequent moves easier and
limits any vibrations.
It is very important to have good air circulation around the instrument during storage. Dust can be
cleaned, but humidity is harmful.
1.9
Laboratory requirements
Ambient temperature:
16 - 30°C, maximum rate of change ± 5°C per hour
Humidity:
20 - 80% relative
Altitude:
2’500 meters above sea level
400 cm min.
35
0
cm
m
in.
min. 80
1
2
3
4
5
6
1 Quantometer ARL 3460
2 Table (for computer and peripherals)
3 Office desk
4 Chair
5 Cupboard (for samples, material, docu-
mentation)
6 Washbasin
Example of laboratory layout