ARL 3460 Metals Analyzer
Page 8/17
Technical Description
2
Max. degree of vacuum :
1 x 10
-3
mm Hg
Temperature control:
Inside of spectrometer is temperature controlled to
±
0.1
°
at 38
°
for
ambient temperature 16 - 30
°
C
Spark stand:
Argon flushed. Stand with built in ignitor for improved trace analysis.
Included closed water cooled system.
The safety circuit protects the operator and the instrument. It is not
possible to perform an analysis when the stand door is open. This
circuit conforms with the international norm CEI 1010-1.
Fatigue lamps:
With built-in intensity level selectable by software. The fatigue
improves the stability of the photomultipliers as well as their life time.
Source excitation model:
HIREP (HIgh REPetition source)
Power :
400 VP
Ignitor :
15 KVP
Spark discharge rate: 400 discharges / sec (400 Hz)
Measuring method:
Integration
Spectrometer control:
ARL MMB 88 microprocessor utilising CMOS technology with status
measuring card A/D converters and attenuators included for each
channel. Up to 60 analytical channels
Dust protection:
Built-in dust protection with high capacity cooling fans
Argon Flow:
3.5
L/min during analysis
0
.
3 L /min stand-by
Dimension:
Height: 1
19 c
m
Width: 16
6
.5 cm
Depth:
91
cm
91
85.5
1
19
16
6
.5
13
6
.5
30
67
Approximate weight:
Net 4
5
0 Kg approximately
Packing
Sealed airfreight packing as follows:
Packing
Material
Gross
weight
Dimension
(cm)
Volume
m
3
1 case
ARL 3460
748
195 x 140 x 145
3,958
1 case
Screen
27
55 x 53 x 49
0.142