Installation
Page 90
Integral Fraction Collector FT (VF-F20-A) Operating Manual
6. Check the flush solvent tubing across the entire flow path:
a) Make sure the tubing is not bent, pinched or squeezed at any
point in the flow path.
b) If you have to cut tubing to length, use a tubing cutter. Make
sure that the cut is at right angle to the length of the line.
Purging the Flush Pump
After the device has been turned on, purge the flush pump using
Chromeleon to fill the flush pump with the fresh flush solvent.
During purging, the flush pump and flush buffer loop are washed
continuously until all flush purge lines are filled with the fresh flush
solvent.
5.9.3
Connecting the Delay Capillary
Parts required
Delay Capillary
Follow these steps
1. Remove the detector waste line if it has been installed onto detector
flow cell outlet or the outlet (OUT) on the flow cell connection unit.
Figure 36: Removing the detector waste line
No.
Description
1
Detector waste line
Summary of Contents for VF-F20-A
Page 4: ...Contacting Us Page 4 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 17: ...1 Using this Manual Integral Fraction Collector FT VF F20 A Operating Manual Page 17 ...
Page 31: ...2 Safety Integral Fraction Collector FT VF F20 A Operating Manual Page 31 ...
Page 48: ...3 Device Overview Page 48 Integral Fraction Collector FT VF F20 A Operating Manual ...
Page 215: ...8 Troubleshooting Integral Fraction Collector FT VF F20 A Operating Manual Page 215 ...
Page 221: ...9 Specifications Integral Fraction Collector FT VF F20 A Operating Manual Page 221 ...
Page 236: ...12 Index Page 236 Integral Fraction Collector FT VF F20 A Operating Manual 12 Index ...