9
•
Specifications
Page 74
Vanquish Charger Operating Manual
9.1
Performance Specifications
The Charger performance is specified as follows:
Type
Specification
Temperature range
4-40 °C (settable),
cooling target
≥
22 K below ambient
Temperature accuracy
-2 °C / +4 °C
Spatial temperature deviation
±2 °C
Plate capacity
9 deep well plates, maximum height 45 mm or
9 sample racks, maximum height 45 mm including vials or
20 well plates, maximum height 20 mm
Sample capacity
Up to 7,680 samples in twenty 384 well plates
Cycle time for handling system
< 1 min (shuttling a sample rack or well plate from or to autosampler)
PC connection
USB 2.0 interface
Control
Chromeleon 7, Chromeleon 6.8
The Charger can be operated also with other data systems. For details,
contact the Thermo Fisher Scientific sales organization.
Safety features
Sample rack/well plate detection sensor on shovel,
Barcode reader to detect rack type and shelf ID
Automatic stop when door is opened (interlock door switch)
Automatic opening of door after power failure
3 LEDs for status monitoring
Good Laboratory Practice (GLP)
features
All system parameters are logged in the Chromeleon Audit Trail.
Summary of Contents for Vanquish Charger
Page 1: ...Vanquish Charger Operating Manual 4829 0001 EN Revision 1 0a September 2017 thermoscientific ...
Page 4: ...Contacting Us Page 4 Vanquish Charger Operating Manual ...
Page 8: ...Contents Page 8 Vanquish Charger Operating Manual ...
Page 66: ...7 Maintenance Page 66 Vanquish Charger Operating Manual ...
Page 76: ...9 Specifications Page 76 Vanquish Charger Operating Manual ...
Page 80: ...10 Accessories Consumables and Replacement Parts Page 80 Vanquish Charger Operating Manual ...
Page 85: ...12 Index Vanquish Charger Operating Manual Page 85 12 Index ...