6
•
Operation
Page 50
Vanquish Charger Operating Manual
Placing Sample Racks or Well Plates in the Shelf
A barcode reader allows rack type identification of Vanquish sample
racks and well plates with a barcode for rack type identification as
shown in
Figure 7: Sample rack with barcodes for rack type identification
CAUTION—Positioning of Sample Racks or Well Plates without Barcode
Sample racks and well plates without a special Vanquish
barcode may
easily be placed the wrong way around, but Chromeleon cannot detect
the incorrect orientation. This may result in a safety hazard.
•
Preferably, use Vanquish sample racks and well plates with a
barcode for rack type identification, as this allows the software to
check the correct orientation. Refer to the current list of sample
racks and well plates approved by Thermo Fisher Scientific.
•
When you place sample racks and well plates without such barcodes,
double-check the correct positioning as described in this section.