1
Rotor Specifications
Sorvall
Thermo Scientific
75003607 TX-750
1-7
Table 1-6.
Sorvall 230V 50Hz/60Hz rectangular bucket 75003614
Centrifuge
Sorvall Legend XT
Sorvall Legend XF
Sorvall ST 40
Catalog #
75004505
75004532
75004509
Weight empty [kg]
6,8
6,8
6,8
max. cycle number
50,000
50,000
50,000
Maximum permissible load [g]
4x750
4x750
4x750
maximum speed n
max
[rpm]
4,500
4,500
4,500
maximum RCF value at n
max
4,415
4,415
4,415
max. / min. radius [cm]
19.5 / 8.9
19.5 / 8.9
19.5 / 8.9
Pitch angle [°]
90
90
90
Accel. / braking time [s]
45 / 50
45 / 50
45 / 50
sample heating at n
max
[° C] referred to ambient
temperature of 23 °C, running time 120
minutes
11
11
11
aerosol-tight
*
*
Tested by HPA, Porton-down, UK
yes
yes
yes
permissible temperature range autoclavable
°C
121
121
121
Centrifuge
Sorvall Legend XTR
Sorvall Legend XFR
Sorvall ST 40R
Catalog #
75004520
75004538
75004524
Weight empty [kg]
6,8
6,8
6,8
max. cycle number
50,000
50,000
50,000
Maximum permissible load [g]
4x750
4x750
4x750
maximum speed n
max
[rpm]
4,500
4,500
4,500
maximum RCF value at n
max
4,415
4,415
4,415
max. / min. radius [cm]
19.5 / 8.9
19.5 / 8.9
19.5 / 8.9
Pitch angle [°]
90
90
90
Accel. / braking time [s]
40 / 50
40 / 50
40 / 50
aerosol-tight
*
yes
yes
yes
permissible temperature range autoclavable
°C
121
121
121
*
Tested by HPA, Porton-down, UK