Connecting the Sample Flow to the Source
Connect the sample flow to the H-ESI source or the APCI source.
• Connecting the Sample Flow to the H-ESI Source
• Connecting the Sample Flow to the APCI Source
Connecting the Sample Flow to the H-ESI Source
Connect the H-ESI source or the low-flow H-ESI source to the sample flow. After making the
connections check for leaks at each of the connections before you turn the power on.
Note
Use the fittings (PEEK or Viper™) that are appropriate for your application. The
illustrations are examples only.
Note
Refer to the
OptaMax NG Ion Source User Guide
for more information.
CAUTION
If liquid is leaking out of a tube or a fitting it may be electrically live and could
cause an electric shock when touched. Turn off the power to the source before fixing the
leak. Turn the power off by doing one of the following:
• Switch the electronics service switch to service mode (down position) (1 in
-or-
• Switch the main power switch to off mode (down position) (5 in
turn off the power to the source, the MS and the forepumps.