1
Getting Familiar with Your TriPlus RSH
SPME Conditioning Station
Thermo Scientific
TriPlus RSH User Guide
61
The
total area
is determined from the MHE-plot of
Consequently, the acetone content can be determined as follows:
1. Calculate the total area using the Excel spreadsheet. In this example the total area is 101.1
2. Compare the total area against the calibration plot from which the amount can be
determined as 19 μg. See
.
SPME Conditioning Station
The SPME Conditioning Station is an optional device used for the SPME (Solid Phase Micro
Extraction) technique. See
It conditions or cleans a fiber between chromatographic runs.
A flow of inert gas protects the fiber from degradation when it is exposed to elevated
temperatures. The temperature for conditioning or cleaning can be selected.
Figure 44.
SPME Conditioning Station
A spare fiber can be conditioned in the front well to ensure that a spare fiber is always ready
for use. The front position is designed to condition a spare SPME Fiber. See
WARNING
The SPME Conditioning Station operates at a maximum temperature of
350 °C. Operating at this temperature, the outer surface will reach approximately 70 °C.
Do not touch the outside of the SPME Conditioning Station.